Description

Combination Powder and Fume Hood with HEPA filtration and house exhaust connection designed to process liquid and powder substances. Fume Hood designed to house stir plate and analytical balance. It features cup sink; air, water, and gas fittings; and alarm for monitoring and maintaining recommended face velocity of 75 LFPM. Fume and Powder Hoods separated by sliding door to eliminate cross-contamination during processing. Powder Hood uses front airfoils and rear plenums to maintain laminar airflow across the work surface, which is vented through a HEPA filtration system to house exhaust.

Size:
76″ Exterior Width
33″ Exterior Depth
39″ Interior Height

FSI Category: 8-9

Videos

 

 

About

Contained Vented Enclosures (CVEs) – Laminar Flow Hoods

Single HEPA Filter – Bag Out Series

Flow Sciences’ Single HEPA Series Contained Vented Enclosures are designed for maximum containment with the use of airfoils and plenums, creating a laminar airflow across the work surface, which reduces eddies and turbulence. The ETA Series features a front door lift that allows for easy loading and unloading of equipment. The single 4” thick HEPA filter and fan is ideal for benchtop / scale up and nanotechnology applications, and has been proven by an independent test organization to contain loss to below 1 µg/m^3 based on process and quality. The top-mounted exhaust system can recirculate air into the lab or send it directly to the house exhaust without transferring vibration from the fan unit to the work surface. As an additional safety measure, FSI’s Contained Vented Enclosures feature a face velocity alarm that visually and audibly alerts the operator when airflow is compromised. These features allow for reproducibility and accuracy of weighing. In case of an accident, the dished phenolic base is chemically resistant and designed to contain spills.  The ETA Series is available in 3’ or 4’ standard width units and features options such as LED lights and waste ports to safely transfer contaminants from the enclosure.

Dual HEPA Filters – Bag In / Bag Out Series

Flow Sciences’ Dual HEPA BIBO Series of Contained Vented Enclosures are designed to prevent loss of containment with the use of airfoils and plenums, creating a laminar airflow across the work surface, which reduces eddies and turbulence. The ETA Series features a front door lift that allows for easy loading and unloading of equipment. The single 4” thick HEPA filter with fan is ideal for benchtop / scale up and nanotechnology applications, and has been proven by an independent test organization to contain loss to below 1 µg/m^3 based on process and quality. FSI’s state-of-the-art BIBO feature allows the operator to safely remove and replace the main filter while maintaining maximum containment from toxic fumes and particles. The top-mounted exhaust system can recirculate air into the lab or send it directly to the house exhaust without transferring vibration from the fan unit to the work surface. As an additional safety measure, FSI’s Contained Vented Enclosures feature a face velocity alarm that visually and audibly alerts the operator when airflow is compromised. These features allow for reproducibility and accuracy of weighing. In case of an accident, the dished phenolic base is chemically resistant and designed to contain spills. The ETA Series is available in 3’ or 4’ standard width units and features options such as LED lights and waste ports to safely transfer contaminants from the enclosure.

Details

CLASSIFICATION

CLASSIFICATION

CFM REQUIREMENTS

CFM REQUIREMENTS

VOLTAGE / FILTER OPTIONS

VOLTAGE / FILTER OPTIONS


Features

OVERLAPPING SASH BYPASS PROVIDES BETTER CONTAINMENT AND DOES NOT REQUIRE CHANGES IF VAV IS INSTALLED IN THE LAB FUME HOOD
 

SAVE OVER 60% OF ENERGY WITH THE SAF T FLOW FUME HOOD SERIES OVER OTHER LAB FUME HOODS

GREAT ANSI/ASHRAE 110 CONTAINMENT RESULTS UNDER A WIDE RANGE OF CONTSTANT VOLUME AND VAV CONDITIONS IN THE LOW FLOW SAF T FLOW CHEMICAL FUME HOOD SERIES

DESIGN. Uniquely designed for safety and convenience, the SAF T FLOW™ features include a different kind of bypass system that assures all airflow entrained by the hood is actively used to engage, contain, and exhaust fumes from the containment area. High efficiency allows a high degree of energy saving which saves lots of money.

SAFETY. The SAF T FLOW™ lab fume hood series provides safety from chemical vapors generated during chemical processes including reflux, distillation, flash chromatography, evaporation, and other process equipment applications.
 
FLEXIBILITY. SAF T FLOW™ chemical hood uses the same overall design structure in all standard fume hood sizes to maximize simplicity in lab planning, layout, and HVAC coordination. This approach allows any popular low flow exhaust option without requiring expensive retrofits to the hood.

SLIDE-OUT TOP SYSTEM.  Sometimes tops shatter. The slide-out feauture allows complete top replacement without lifting the fume hood superstructure and disconnecting plumbing and ductwork.

STABILIZING BAFFLE CONTROL SYSTEM.  Slots orented in both X and Y axis promote even air flow back into the hood interior from top to bottom of the sash opening. The rear baffle is also removable for easy fume hood cleaning.
 
QUICK-SET SCAFFOLDING SYSTEM.  Detachment knobs make for quick scaffolding setup and easy rear baffle removal for cleaning of the fume hood.
 
CLEAR CORD FLIP-UP AIRFOIL.  Used for safe cord routing and airflow direction in the chemical fume hood.
 
HEAVY GUAGE SUPERSTRUCTURE AND CHAIN DRIVE.  14 and 18 guage framing system and #35 steel chain drive will run 500,000 cycles without breaking.

OVERLAPPING SASH BYPASS

OVERLAPPING SASH BYPASS

The overlapping sash bypass provides better containment and never needs to be removed or replaced if Variable Air Volume conservation systems are retrofitted into an existing Saf T Flow installation. In all exhaust applications, bypass air does productive work clearing fumes from behind the sash and means all exhausted air does containment work at all times, with any exhaust strategy.
 
 
Keep in mind, volumetric reductions in fume hood exhaust can save thousands of energy dollars each year on each hood. This is the main reason traditional constant volume hoods have been frequently converted to VAV to save exhaust air and energy.
 
We have tested our OSB from 100 FPM full open sash down to 60 FPM at 18” sash opening with ASHRAE 110 containment equipment and cannot tell any difference in containment performance at any selected face velocity or opening within this range. Breathing zone containment ranged from 0.005ppm down to undetectable. Fantastic containment at very low exhaust volumes! 

Sustainability


When Considering Sustainability,
Which Fume Hood is Really the Best Deal?


In the present cost-conscious economy, there is substantial pressure on building contractors to buy the least expensive exhaust hood and focus on sustainability. What is saved on such inexpensive hoods at the outset is often sacrificed over time in higher maintenance and energy costs. The true cost of a fume hood is made up of the following:

1) The Original Cost to Purchase & Install

2) The Yearly Maintenance Cost

3) The Yearly Energy Expense

Since building contractors typically only focus on hood costs; maintenance and energy costs have become the problem of someone else. Fume hood cost of ownership should include an analysis of all three costs to reach a sensible overall purchase and construction plan for the lab owner. If cost, maintenance, and energy consumption are all minimized, the final hood choice will also be the most sustainable.


COST

OF

OWNERSHIP

Here’s why all three costs are relevant. FSI and Wave Consulting evaluated the SAF T FLOW™ fume hood and ten competitors for purchase price, maintenance cost, and energy consumption. The initial purchase price of the FSI unit was more expensive, but by comparing the energy usage and maintenance costs of the SAF T FLOW™ unit versus Brand G and Brand I, we can clearly see that after 15 years (an often-used construction cycle time for labs and leased equipment), the FSI hood is the most cost efficient option.



Exhausting Conditioned Air and
Maintenance Costs are Very Expensive.

Obviously, any lab owner should recognize that the initial costs of containment systems are not the only price that should be taken into account during the hood selection process. The technology used in the SAF T FLOW™ Fume Hood series requires less CFM and still maintains excellent containment. This results in significant yearly savings, and drastic savings over a 15 year lab cycle time. The SAF T FLOW™ Fume Hood performs better and more efficiently and includes advanced design features which cost more on the front end. This extra initial cost is rapidly recovered; usually in the second year of service.



When you puchase a SAF T FLOW™ Fume Hood, you will see the break even point in 1.8 years compared to legacy hoods that require more energy and require more maintanance. Total cost of ownership is the only way to evaluate any purchasing decision. Many companies are considering the equipment purchase based on cost of ownership, and you should too.



Learn more about the sustainability of the SAF T FLOW™ Fume Hood series by contacting an FSI containment expert and asking for the SAF T FLOW™ Cost of Ownership white paper. There you will see these statistics and more!



Options

SERIES MODELS

SERIES MODELS

STANDARD OPTIONS

STANDARD OPTIONS

PLUMBING OPTIONS

PLUMBING OPTIONS

Exhaust Options

FAN / FILTER
DETAILS AND
DIMENSIONS

NO FAN / WITH FILTER
DETAILS AND
DIMENSIONS

NO FAN /NO FILTER
DETAILS AND
DIMENSIONS

NO FAN / BIBO
DETAILS AND
DIMENSIONS

 FAN / FILTER
DETAILS AND
DIMENSIONS

MESH FAN / BIBO
DETAILS AND
DIMENSIONS

MESH FAN / INVERTED
DETAILS AND
DIMENSIONS


Exhaust Option A

FAN / FILTER

The most common exhaust option for pharmaceutical compunding is our fan / filter option. With a 4″ pleated HEPA filter and fan, this enclosure is ideal for working with APIs and many more applications. The top mount fan has a 6″ exhaust port that allows for circulation into lab or connection to house exhaust. FSI has a bag out filter procedure to replace this filter after it’s 3-5 year life span.

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA242424AAA23.75"
(60.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
42.48"
(107.8 cm)
109 lb
(49 kg)
ETA362424AAA35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
138 lb
(63 kg)
ETA482424AAA47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
173 lb
(78 kg)
ETA243024AAA23.75"
(60.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.56"
(108.1 cm)
121 lb
(55 kg)
ETA363024AAA35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.51"
(108.1 cm)
152 lb
(69 kg)
ETA483024AAA47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
44.31"
(112.5 cm)
193 lb
(88 kg)
ETA603027AAA59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
274 lb
(124 kg)
ETA723027AAA71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
302 lb
(137 kg)
ETA963027AAA95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
382 lb
(173 kg)

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA242424AAD23.75"
(60.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
42.48"
(107.8 cm)
109 lb
(49 kg)
ETA362424AAD35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
138 lb
(63 kg)
ETA482424AAD47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
173 lb
(78 kg)
ETA243024AAD23.75"
(60.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.56"
(108.1 cm)
121 lb
(55 kg)
ETA363024AAD35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.51"
(108.1 cm)
152 lb
(69 kg)
ETA483024AAD47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
44.31"
(112.5 cm)
193 lb
(88 kg)
ETA603027AAD59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
274 lb
(124 kg)
ETA723027AAD71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
302 lb
(137 kg)
ETA963027AAD95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
382 lb
(173 kg)

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA242424AAC23.75"
(60.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
42.48"
(107.8 cm)
109 lb
(49 kg)
ETA362424AAC35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
138 lb
(63 kg)
ETA482424AAC47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
173 lb
(78 kg)
ETA243024AAC23.75"
(60.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.56"
(108.1 cm)
121 lb
(55 kg)
ETA363024AAC35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.51"
(108.1 cm)
152 lb
(69 kg)
ETA483024AAC47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
44.31"
(112.5 cm)
193 lb
(88 kg)
ETA603027AAC59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
274 lb
(124 kg)
ETA723027AAC71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
302 lb
(137 kg)
ETA963027AAC95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
382 lb
(173 kg)

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA242424AAF23.75"
(60.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
42.48"
(107.8 cm)
109 lb
(49 kg)
ETA362424AAF35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
138 lb
(63 kg)
ETA482424AAF47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
44.31"
(112.5 cm)
173 lb
(78 kg)
ETA243024AAF23.75"
(60.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.56"
(108.1 cm)
121 lb
(55 kg)
ETA363024AAF35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
42.51"
(108.1 cm)
152 lb
(69 kg)
ETA483024AAF47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
44.31"
(112.5 cm)
193 lb
(88 kg)
ETA603027AAF59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
274 lb
(124 kg)
ETA723027AAF71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
302 lb
(137 kg)
ETA963027AAF95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
46.93"
(119.2 cm)
382 lb
(173 kg)


Exhaust Option B

FAN / BIBO

The bag in / bag out (BIBO) filter cahnge method is unique to FSI and we are very proud to provide double HEPA filtration and ease of filter change. This system allows for circulation into the lab and also connection straight to house exhaust. The BIBO system allows for the filter to be changed in a clean environment where the saturated filter is not exposed to the lab environment.

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA362424ABA35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
55.31"
(135.4 cm)
182 lb
(83 kg)
ETA482424ABA47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
53.23"
(135.1 cm)
231 lb
(105 kg)
ETA363024ABA35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
51.56"
(130.9 cm)
206 lb
(93 kg)
ETA483024ABA47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
53.23""
(135.1 cm)
251 lb
(112 kg)
ETA603027ABA59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
360 lb
(163 kg)
ETA723027ABA71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
388 lb
(176 kg)
ETA963027ABA95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.81"
(141.9 cm)
498 lb
(226 kg)

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA362424ABD35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
55.31"
(135.4 cm)
182 lb
(83 kg)
ETA482424ABD47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
53.23"
(135.1 cm)
231 lb
(105 kg)
ETA363024ABD35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
51.56"
(130.9 cm)
206 lb
(93 kg)
ETA483024ABD47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
53.23""
(135.1 cm)
251 lb
(112 kg)
ETA603027ABD59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
360 lb
(163 kg)
ETA723027ABD71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
388 lb
(176 kg)
ETA963027ABD95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.81"
(141.9 cm)
498 lb
(226 kg)

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA362424ABC35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
55.31"
(135.4 cm)
182 lb
(83 kg)
ETA482424ABC47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
53.23"
(135.1 cm)
231 lb
(105 kg)
ETA363024ABC35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
51.56"
(130.9 cm)
206 lb
(93 kg)
ETA483024ABC47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
53.23""
(135.1 cm)
251 lb
(112 kg)
ETA603027ABC59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
360 lb
(163 kg)
ETA723027ABC71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
388 lb
(176 kg)
ETA963027ABC95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.81"
(141.9 cm)
498 lb
(226 kg)

MODELEXT. WIDTHEXT. DEPTHINT. HEIGHTINT. DEPTHTOT. HEIGHTWEIGHT
ETA362424ABF35.75"
(90.8 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
55.31"
(135.4 cm)
182 lb
(83 kg)
ETA482424ABF47.75"
(121.3 cm)
24"
(61 cm)
24"
(61 cm)
20.57"
(52.2 cm)
53.23"
(135.1 cm)
231 lb
(105 kg)
ETA363024ABF35.75"
(90.8 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
51.56"
(130.9 cm)
206 lb
(93 kg)
ETA483024ABF47.75"
(121.3 cm)
30"
(76.2 cm)
24"
(61 cm)
26.55"
(67.4 cm)
53.23""
(135.1 cm)
251 lb
(112 kg)
ETA603027ABF59.75"
(151.8 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
360 lb
(163 kg)
ETA723027ABF71.75"
(182.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.93"
(142.1 cm)
388 lb
(176 kg)
ETA963027ABF95.75"
(243.2 cm)
30"
(76.2 cm)
27"
(68.6 cm)
26.55"
(67.4 cm)
55.81"
(141.9 cm)
498 lb
(226 kg)


Exhaust Option C

NO FAN / WITH FILTER

This enclosure is designed for applications that need to be filtered but still go to house exhaust. Featuring a 4″ HEPA filter and exhaust cap, a 6″ exhaust port should be attached to existing house exhaust. This is enclosure can also be used for explosion proof (XP) applications.


Exhaust Option D

NO FAN / NO FILTER

This exhaust option is the most similar to the original VBSE™ from Flow Sciences. With no fan and no filter, this enclosure is versitile and can be used to connect directly to house exhaust or to a remote fan/filter. Labs with upper cabinets could benefit from this enclosure as the total height is lesser than any other exhaust option.


Exhaust Option E

NO FAN / BIBO

This exhaust option is engineered for applications that need dual HEPA filtration and also sent directly to house exhaust. The Bag in / Bag out (BIBO) filter change procedure includes two 4″ HEPA filters with a simple primary filter changeout method that never exposes the lab to the saturated filter. This enclosure is also used for explosion proof (XP) applications.


Exhaust Option F

MESH FAN / FILTER


Exhaust Option G

MESH FAN / BIBO


Exhaust Option H

MESH FAN / INVERTED

Get a Quote / Info

  • What is being done inside of the enclosure? What type of material (powder, liquid, gas, nuisance odor) is being worked with? How does the material enter and exit the enclosure system? etc...
  • What type of filtration is required? Single HEPA, Dual HEPA, Carbon, House Exhaust, etc... What is the required OEL (Occupational Exposure Limit) for the process, or any other details about containment goals? What is the quantity of powder or liquid, task duration, composition of powder, etc...?
  • What equipment is being worked with? What is the equipment model, size, scope, function, and any other information that will affect the design of the enclosure, including movement, heat output, etc...? *State the specific equipment make and model if available*
  • Drop files here or
  • Are there any additional notes or information that should be considered? Are there any special design requirements?

CFM, Static Pressure

CFM, Duct Static Pres.  @ Face Velocities at a Glance

Metric Conversions:

1CFM = 1.70 CM/H

1” Static Pressure = 249Pa

1Foot = 305 mm

CS=Combo Sash: a single combo sash ½ open with verticals closed

NA= This Sash Setting not recommended

 2025 Mercantile Dr., Leland, NC 28451
 Phone: 1-800-849-3429
 Fax: (910) 763-1220

© 2019 Flow Sciences, Inc. All rights reserved